Removal of Air Emboli In Vortex Head
If the venous reservoir completely empties of volume, the venous return
is zero and air has entered the vortex head:
- Clamp the arterial line and come off bypass.
- Inform the Surgeon you have come off bypass.
- Establish the problem with venous drainage, once drainage has returned
the vortex head must be reprimed.
- Clamp the arterial outlet of the oxygenator.
- Remove the vortex head from the console, lower the head and remove
the air.
- Clamp the vortex head outlet.
- Open the recirculation line.
- Replace the vortex head on the console, increase the pump speed.
- Release clamp at the vortex head outlet, removing the air.
- Circulate for a couple of minutes until all the air has been removed
and purged through the recirculation line.
- Remove clamp from the arterial outlet of the oxygenator.
- Initiate bypass again.
- Increase Fi02
- Increase arterial mean pressure, open recirculation line and maintain
a high arterial mean pressure.
- Note period off bypass.
- Send arterial blood sample to check blood gas status.
|